MICRO-EPSILON TIM 160 Manual de Instruções Página 20

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 64
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 19
Page 20
Optical Charts
thermoIMAGER TIM
TIM 160/200
(160 x 120 px)
Focal
length
Minimum
distance
1
Distance to object (Measuring field in m, pixel in mm)
0.02 0.1 0.2 0.3 0.5 1.2 2 4 6 10 30 100
72 ° x 52 °
Wide angle
3.3 mm 0.2 m HFOV
(m)
0.029 0.15 0.29 0.44 0.73 1.45 2.91 5.81 8.72 14.5 43.6 145.3
VFOV
(m)
0.020 0.10 0.20 0.29 0.49 0.98 1.95 3.90 5.85 9.8 29.3 97.5
IFOV
(mm)
0.182 0.91 1.82 2.72 4.54 9.08 18.16 36.33 54.5 90.8 272.5 908.2
Fig. 3 Table with examples (TIM 160 / 200) showing what spot sizes and pixel sizes will be reached in which distance
HFOV: Horizontal enlargement of the total measuring field at object level
VFOV: Vertical enlargement of the total measuring field at object level
IFOV: Size of the single pixel at object level
1) The accuracy of measurement can be outside of the specifications for distances below the defined minimum distance.
Fig. 4 Measurement field of the thermoIMAGER TIM representing the standard lens 23 ° x 17 °
Dimensions in mm, not to scale
Vista de página 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 63 64

Comentários a estes Manuais

Sem comentários